Background

Indications

Contraindications

Procedure

True electrical capture
False capture with visible phantom beats[2]
  • Pad placement:
    • Pad on apex of heart and on right upper chest
    • Pad on lead V3 position and between left scapula and T-spine
  • Set: HR 80, pacing threshold usually between 40-80 mA
    • If hemodynamically unstable or with evidence of end organ poor perfusion start at 80 mA and titrate down as tolerated.
    • Look for clear QRS complex and T-wave following pacer spike
    • Check pulse to confirm mechanical capture
    • Final current set 5-10 mA above threshold level for patient

See Also

External Links

References

  1. 1.0 1.1 Epstein AE, DiMarco JP, Ellenbogen KA, et al. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices): developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons. Circulation. May 27 2008;117(21):e350-408