Tono-Pen use: Difference between revisions
(Created page with "==CALIBRATION== The Pen should be calibrated if the batteries have been changed or if the LCD reads “CAL” followed by “- - - -”. Remember, calibration involves depress...") |
No edit summary |
||
| (8 intermediate revisions by 2 users not shown) | |||
| Line 1: | Line 1: | ||
== | ==To Use (if already calibrated)== | ||
*Depressing the Operator Button ONCE | |||
*Monitor should show TWO lines (====). | |||
*Use on eye | |||
TWO lines (====). | |||
==To Calibrate== | |||
*The Pen should be calibrated if the batteries have been changed or if the LCD reads “CAL” followed by “- - - -”. | |||
*Calibration involves depressing the Operator Button TWICE and getting ONE line (- - - -). | |||
*If so, | |||
*#Point the transducer end straight down towards the floor. | |||
*#Press the Operator Button “distinctly, two times, within 1.5 seconds.” | |||
*#Initially “CAL” appears on the LCD, followed by “UP” (after about 15 - 30 seconds!) | |||
*#Invert the transducer end to point straight up toward the sky. | |||
*#*A properly functioning Pen will display “”Good”. | |||
*#*If “bAd”, repeat the calibration again. | |||
===If unsuccessful=== | |||
Options include: | |||
*clean the tip (compressed air is recommended) | |||
*change the batteries. | |||
**Use the little probe that came with the Pen to open the battery compartment. | |||
*Tech Support | |||
**www.reichert.com | |||
**www.tonopen.us | |||
**1-888-849-8955 | |||
==See Also== | |||
*[[Eye pain]] | |||
*[[Acute angle closure glaucoma]] | |||
*[[IOP]] | |||
[[Category:Procedures]] | [[Category:Procedures]] | ||
[[Category:Ophthalmology]] | |||
==External Links== | |||
===Videos=== | |||
{{#widget:YouTube|id=sBosdQ6o1oY}} | |||
Latest revision as of 21:52, 29 June 2021
To Use (if already calibrated)
- Depressing the Operator Button ONCE
- Monitor should show TWO lines (====).
- Use on eye
To Calibrate
- The Pen should be calibrated if the batteries have been changed or if the LCD reads “CAL” followed by “- - - -”.
- Calibration involves depressing the Operator Button TWICE and getting ONE line (- - - -).
- If so,
- Point the transducer end straight down towards the floor.
- Press the Operator Button “distinctly, two times, within 1.5 seconds.”
- Initially “CAL” appears on the LCD, followed by “UP” (after about 15 - 30 seconds!)
- Invert the transducer end to point straight up toward the sky.
- A properly functioning Pen will display “”Good”.
- If “bAd”, repeat the calibration again.
If unsuccessful
Options include:
- clean the tip (compressed air is recommended)
- change the batteries.
- Use the little probe that came with the Pen to open the battery compartment.
- Tech Support
- www.reichert.com
- www.tonopen.us
- 1-888-849-8955