Deterioration after intubation: Difference between revisions
Ostermayer (talk | contribs) (Text replacement - " pts" to " patients") |
Neil.m.young (talk | contribs) (Text replacement - " pt " to " patient ") |
||
| Line 20: | Line 20: | ||
**Attach end-tidal CO2 to verify and check depth (cm at lip) | **Attach end-tidal CO2 to verify and check depth (cm at lip) | ||
*O - Obstruction of tube/circuit | *O - Obstruction of tube/circuit | ||
**Use suction catheter to remove mucus plug, or make sure | **Use suction catheter to remove mucus plug, or make sure patient not biting down | ||
*P - Pneumothorax | *P - Pneumothorax | ||
**Verify via [[Ultrasound: Lungs|ultrasound]], CXR, or [[needle thoracostomy]] (high suspicion) | **Verify via [[Ultrasound: Lungs|ultrasound]], CXR, or [[needle thoracostomy]] (high suspicion) | ||
| Line 29: | Line 29: | ||
===Fix=== | ===Fix=== | ||
*Disconnect vent and put light pressure on | *Disconnect vent and put light pressure on patient chest | ||
*Oxygen 100% | *Oxygen 100% | ||
**Bag and take time to evaluate your patient | **Bag and take time to evaluate your patient | ||
Revision as of 18:19, 1 July 2016
Background
Clinical Features
- Desaturation, other vital sign abnormalities, or full arrest while on mechanical ventilation
Differential Diagnosis
DOPE[1][2]
- Displaced ETT
- Obstruction (anywhere along circuit)
- Pneumothorax
- Equipment failure (ventilator malfunction or disconnect)
Diagnosis
- Clinical
Management
Troubleshoot
Immediately disconnect from ventilator (allows for expiration of stacked breaths)
- D - Displacement of tube
- Attach end-tidal CO2 to verify and check depth (cm at lip)
- O - Obstruction of tube/circuit
- Use suction catheter to remove mucus plug, or make sure patient not biting down
- P - Pneumothorax
- Verify via ultrasound, CXR, or needle thoracostomy (high suspicion)
- E - Equipment failure
- Connect to BVM
- S - Stacked breaths - Auto-PEEP especially in COPD/Asthma
- Disconnect from ventilator
Fix
- Disconnect vent and put light pressure on patient chest
- Oxygen 100%
- Bag and take time to evaluate your patient
- Tube Position & Function
- Pass bougie or suction all the way through the tube, OR take a look with DL
- Tweak Vent Settings
- Drop TV, then decrease RR, and then increase flow rate
- Caution as it causes hypercapnia and resp acidosis, which is harmful in patients with increased ICP or tox ingestion
- Sonography and CXR
Auto-PEEP (Breath stacking) troubleshooting options
- Bronchodilators if COPD/asthma
- Decrease RR
- Decrease I:E ratio (increase expiratory time)
- Quicker inspiratory flow rate
- Decrease TV
- Increase sedation
See Also
Mechanical Ventilation Pages
- Noninvasive ventilation
- Intubation
- Mechanical ventilation (main)
- Miscellaneous
References
- ↑ EMRA Critical Care Handbook
- ↑ Monica E. Kleinman et al. Pediatric Advanced Life Support: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. AAP. 2010. http://pediatrics.aappublications.org/content/126/5/e1361.full