Difference between revisions of "Initial mechanical ventilation settings"

(MODES)
(Table)
Line 11: Line 11:
 
#  PS (pressure support) 5-8cm to overcome ET
 
#  PS (pressure support) 5-8cm to overcome ET
  
==Table==
+
== Table ==
DZ TV^        RR          I:E      PEEP
 
  
Nl lung 8   10-12     1:2       4
+
{| cellspacing="1" cellpadding="1" border="1" width="200"
 +
|-
 +
| '''Disease'''
 +
| '''TV^'''
 +
| '''RR'''
 +
| '''I:E'''
 +
| '''PEEP'''
 +
|-
 +
| Nl lung
 +
| 8
 +
| 10-12
 +
| 1:2
 +
| 4
 +
|-
 +
| Obstruct
 +
| 6
 +
| 5-8
 +
| 1:4
 +
| 4
 +
|-
 +
| ARDS
 +
| 6
 +
| 12-20
 +
| 1:2
 +
| 2-15
 +
|-
 +
| Hypovol
 +
| 8
 +
| 10-12
 +
| 1:2
 +
| 0-4
 +
|}
  
Obstruct  6      5-8 1:4    4
 
  
ARDS  6        12-20 1:2  2-15
 
  
Hypovol    8      10-12 1:2  0-4
+
^cc/kg  
 
 
^cc/kg
 
  
 
^^FiO2 always = 1.0
 
^^FiO2 always = 1.0
 
  
 
==Changes==
 
==Changes==

Revision as of 21:14, 22 July 2011

Initial

  1. FiO2 100% and ween down
  2. Rate 8-12/min (consider 5-6 for asthma w/ permissive hypercapnea)
  3. Mode
    1. A/C = defalt (most)
    2. SIMV = with obstructive airway disease and an intact respiratory effort (e.g. some COPD, asthma)
    3. PC = with intact respiratory effort and non-severe respiratory failure (prefered in chronic vent)
  4. Peep 0-5 mmH20
  5. TV 5-8 cc/kg (eg. 500-600cc) (adjust to plateau pressure <35 cmH20)
  6. I/E 1:3
  7. PS (pressure support) 5-8cm to overcome ET

Table

Disease TV^ RR I:E PEEP
Nl lung 8 10-12 1:2 4
Obstruct 6 5-8 1:4 4
ARDS 6 12-20 1:2 2-15
Hypovol 8 10-12 1:2 0-4


^cc/kg

^^FiO2 always = 1.0

Changes

O2

  1. PaO2 (ween to SaO2 >90% & PaO2 >60)
  2. FiO2 (ween to <0.5 if poss)
  3. PEEP (see ARDS schedule; if >10, should have art line)
    1. (rule of 7s: decrease 1 FiO2 = 7 pO2)

CO2

  1. PCO2
    1. TV
    2. Rate
  2. (VE-1 x pCO2-1 = VE-2 x pCO2-2)
    1. VE = minute ventilation (RR x TV)

pH

to increase pH 0.01 --> decrease pCO2 by 1

MISC (normally already set)

  1. Inspiratory flow rate = 60L/min (100L/min with asthma)
  2. Sensitivity = 1-2 cmH2O

See Also

Source

2/8/06 DONALDSON (adapted from emedicine, Lampe)