Tono-Pen use: Difference between revisions

No edit summary
(6 intermediate revisions by 2 users not shown)
Line 1: Line 1:
==CALIBRATION==
==To Use (if already calibrated)==
*The Pen should be calibrated if the batteries have been changed or if the LCD reads “CAL” followed by
*Depressing the Operator Button ONCE
*“- - - -”.
*Monitor should show TWO lines (====).
*Remember, calibration involves depressing the Operator Button TWICE and getting ONE
*Use on eye
*line (- - - -). To take a reading requires depressing the Operator Button ONCE and getting
 
*TWO lines (====).
==To Calibrate==
*1. Point the transducer end straight down towards the floor.
*The Pen should be calibrated if the batteries have been changed or if the LCD reads “CAL” followed by “- - - -”.
*2. Press the Operator Button “distinctly, two times, within 1.5 seconds.”
*Calibration involves depressing the Operator Button TWICE and getting ONE line (- - - -).  
*3. Initially “CAL” appears on the LCD, followed by “UP” (after about 15 - 30 seconds!)
*If so,
*4. Invert the transducer end to point straight up toward the sky.
*#Point the transducer end straight down towards the floor.
*5. A properly functioning Pen will display “”Good”.
*#Press the Operator Button “distinctly, two times, within 1.5 seconds.”
*If “bAd”, repeat the calibration again.
*#Initially “CAL” appears on the LCD, followed by “UP” (after about 15 - 30 seconds!)
*If this is all unsuccessful, options include
*#Invert the transducer end to point straight up toward the sky.
*cleaning the tip (compressed air is recommended)
*#*A properly functioning Pen will display “”Good”.
*change the batteries. Use the little probe that came with the Pen to open the
*#*If “bAd”, repeat the calibration again.
*battery compartment.
 
===If unsuccessful===
Options include:
*clean the tip (compressed air is recommended)
*change the batteries.  
**Use the little probe that came with the Pen to open the battery compartment.
*Tech Support
*Tech Support
*www.reichert.com
**www.reichert.com
*www.tonopen.us
**www.tonopen.us
*1-888-849-8955
**1-888-849-8955


==See Also==
*[[Eye pain]]
*[[Acute angle closure glaucoma]]
*[[IOP]]
[[Category:Procedures]]
[[Category:Procedures]]
[[Category:Ophthalmology]]

Revision as of 15:52, 10 October 2019

To Use (if already calibrated)

  • Depressing the Operator Button ONCE
  • Monitor should show TWO lines (====).
  • Use on eye

To Calibrate

  • The Pen should be calibrated if the batteries have been changed or if the LCD reads “CAL” followed by “- - - -”.
  • Calibration involves depressing the Operator Button TWICE and getting ONE line (- - - -).
  • If so,
    1. Point the transducer end straight down towards the floor.
    2. Press the Operator Button “distinctly, two times, within 1.5 seconds.”
    3. Initially “CAL” appears on the LCD, followed by “UP” (after about 15 - 30 seconds!)
    4. Invert the transducer end to point straight up toward the sky.
      • A properly functioning Pen will display “”Good”.
      • If “bAd”, repeat the calibration again.

If unsuccessful

Options include:

  • clean the tip (compressed air is recommended)
  • change the batteries.
    • Use the little probe that came with the Pen to open the battery compartment.
  • Tech Support
    • www.reichert.com
    • www.tonopen.us
    • 1-888-849-8955