==CALIBRATION==
==CALIBRATION==
The Pen should be calibrated if the batteries have been changed or if the LCD reads “CAL” followed by
*The Pen should be calibrated if the batteries have been changed or if the LCD reads “CAL” followed by
“- - - -”.
*“- - - -”.
Remember, calibration involves depressing the Operator Button TWICE and getting ONE
*Remember, calibration involves depressing the Operator Button TWICE and getting ONE
line (- - - -). To take a reading requires depressing the Operator Button ONCE and getting
*line (- - - -). To take a reading requires depressing the Operator Button ONCE and getting
TWO lines (====).
*TWO lines (====).
1. Point the transducer end straight down towards the floor.
*1. Point the transducer end straight down towards the floor.
2. Press the Operator Button “distinctly, two times, within 1.5 seconds.”
*2. Press the Operator Button “distinctly, two times, within 1.5 seconds.”
3. Initially “CAL” appears on the LCD, followed by “UP” (after about 15 - 30 seconds!)
*3. Initially “CAL” appears on the LCD, followed by “UP” (after about 15 - 30 seconds!)
4. Invert the transducer end to point straight up toward the sky.
*4. Invert the transducer end to point straight up toward the sky.
5. A properly functioning Pen will display “”Good”.
*5. A properly functioning Pen will display “”Good”.
If “bAd”, repeat the calibration again.
*If “bAd”, repeat the calibration again.
If this is all unsuccessful, options include
*If this is all unsuccessful, options include
cleaning the tip (compressed air is recommended)
*cleaning the tip (compressed air is recommended)
change the batteries. Use the little probe that came with the Pen to open the
*change the batteries. Use the little probe that came with the Pen to open the
battery compartment.
*battery compartment.
Tech Support
*Tech Support
www.reichert.com
*www.reichert.com
www.tonopen.us
*www.tonopen.us
1-888-849-8955
*1-888-849-8955


[[Category:Procedures]]
[[Category:Procedures]]

Revision as of 08:47, 19 May 2016

CALIBRATION

  • The Pen should be calibrated if the batteries have been changed or if the LCD reads “CAL” followed by
  • “- - - -”.
  • Remember, calibration involves depressing the Operator Button TWICE and getting ONE
  • line (- - - -). To take a reading requires depressing the Operator Button ONCE and getting
  • TWO lines (====).
  • 1. Point the transducer end straight down towards the floor.
  • 2. Press the Operator Button “distinctly, two times, within 1.5 seconds.”
  • 3. Initially “CAL” appears on the LCD, followed by “UP” (after about 15 - 30 seconds!)
  • 4. Invert the transducer end to point straight up toward the sky.
  • 5. A properly functioning Pen will display “”Good”.
  • If “bAd”, repeat the calibration again.
  • If this is all unsuccessful, options include
  • cleaning the tip (compressed air is recommended)
  • change the batteries. Use the little probe that came with the Pen to open the
  • battery compartment.
  • Tech Support
  • www.reichert.com
  • www.tonopen.us
  • 1-888-849-8955