Difference between revisions of "Tono-Pen use"
(→CALIBRATION) |
|||
Line 1: | Line 1: | ||
− | == | + | ==To Use (if already calibrated== |
− | *The Pen should be calibrated if the batteries have been changed or if the LCD reads “CAL” followed by | + | *Depressing the Operator Button ONCE |
− | + | *Monitor should show TWO lines (====). | |
− | * | + | *Use on eye |
− | + | ||
− | * | + | ==To Calibrate== |
− | * | + | *The Pen should be calibrated if the batteries have been changed or if the LCD reads “CAL” followed by “- - - -”. |
− | * | + | *Calibration involves depressing the Operator Button TWICE and getting ONE line (- - - -). |
− | * | + | *If so, |
− | * | + | *#Point the transducer end straight down towards the floor. |
− | * | + | *#Press the Operator Button “distinctly, two times, within 1.5 seconds.” |
− | *If “bAd”, repeat the calibration again. | + | *#Initially “CAL” appears on the LCD, followed by “UP” (after about 15 - 30 seconds!) |
+ | *#Invert the transducer end to point straight up toward the sky. | ||
+ | *#*A properly functioning Pen will display “”Good”. | ||
+ | *#*If “bAd”, repeat the calibration again. | ||
+ | |||
+ | |||
*If this is all unsuccessful, options include | *If this is all unsuccessful, options include | ||
− | *cleaning the tip (compressed air is recommended) | + | **cleaning the tip (compressed air is recommended) |
− | *change the batteries. Use the little probe that came with the Pen to open the | + | **change the batteries. Use the little probe that came with the Pen to open the |
− | *battery compartment. | + | **battery compartment. |
− | *Tech Support | + | **Tech Support |
− | *www.reichert.com | + | ***www.reichert.com |
− | *www.tonopen.us | + | ***www.tonopen.us |
− | *1-888-849-8955 | + | ***1-888-849-8955 |
[[Category:Procedures]] | [[Category:Procedures]] |
Revision as of 05:31, 11 July 2016
To Use (if already calibrated
- Depressing the Operator Button ONCE
- Monitor should show TWO lines (====).
- Use on eye
To Calibrate
- The Pen should be calibrated if the batteries have been changed or if the LCD reads “CAL” followed by “- - - -”.
- Calibration involves depressing the Operator Button TWICE and getting ONE line (- - - -).
- If so,
- Point the transducer end straight down towards the floor.
- Press the Operator Button “distinctly, two times, within 1.5 seconds.”
- Initially “CAL” appears on the LCD, followed by “UP” (after about 15 - 30 seconds!)
- Invert the transducer end to point straight up toward the sky.
- A properly functioning Pen will display “”Good”.
- If “bAd”, repeat the calibration again.
- If this is all unsuccessful, options include
- cleaning the tip (compressed air is recommended)
- change the batteries. Use the little probe that came with the Pen to open the
- battery compartment.