Difference between revisions of "Tono-Pen use"
(→To Use (if already calibrated) |
(→To Calibrate) |
||
Line 15: | Line 15: | ||
*#*If “bAd”, repeat the calibration again. | *#*If “bAd”, repeat the calibration again. | ||
− | + | ===If unsuccessful=== | |
− | + | Options include: | |
− | * | + | *clean the tip (compressed air is recommended) |
− | + | *change the batteries. | |
− | + | **Use the little probe that came with the Pen to open the battery compartment. | |
− | + | *Tech Support | |
− | + | **www.reichert.com | |
− | + | **www.tonopen.us | |
− | + | **1-888-849-8955 | |
==See Also== | ==See Also== | ||
[[Category:Procedures]] | [[Category:Procedures]] |
Revision as of 05:32, 11 July 2016
To Use (if already calibrated)
- Depressing the Operator Button ONCE
- Monitor should show TWO lines (====).
- Use on eye
To Calibrate
- The Pen should be calibrated if the batteries have been changed or if the LCD reads “CAL” followed by “- - - -”.
- Calibration involves depressing the Operator Button TWICE and getting ONE line (- - - -).
- If so,
- Point the transducer end straight down towards the floor.
- Press the Operator Button “distinctly, two times, within 1.5 seconds.”
- Initially “CAL” appears on the LCD, followed by “UP” (after about 15 - 30 seconds!)
- Invert the transducer end to point straight up toward the sky.
- A properly functioning Pen will display “”Good”.
- If “bAd”, repeat the calibration again.
If unsuccessful
Options include:
- clean the tip (compressed air is recommended)
- change the batteries.
- Use the little probe that came with the Pen to open the battery compartment.
- Tech Support
- www.reichert.com
- www.tonopen.us
- 1-888-849-8955